Hybrid cardiac SPECT/64-slice CTA-derived LV function parameters: Correlation and reproducibility assessment

  • Sobhi Abadi

      Affiliations

    • Medical Imaging Department, Toronto General Hospital, Toronto, Ontario, M5G2C4 Canada
    • Corresponding Author InformationCorresponding author. Tel.: +1 647 2621419.
  • ,
  • Olga R. Brook

      Affiliations

    • Medical Imaging Department, Rambam Health Care Campus, POB 9602, Haifa 31096, Israel
  • ,
  • Shmuel Rispler

      Affiliations

    • Nuclear Medicine Department, Rambam Health Care Campus, POB 9602, Haifa 31096, Israel
  • ,
  • Alex Frenkel

      Affiliations

    • Nuclear Medicine Department, Rambam Health Care Campus, POB 9602, Haifa 31096, Israel
  • ,
  • Ahuva Engel

      Affiliations

    • Medical Imaging Department, Rambam Health Care Campus, POB 9602, Haifa 31096, Israel
  • ,
  • Zohar Keidar

      Affiliations

    • Nuclear Medicine Department, Rambam Health Care Campus, POB 9602, Haifa 31096, Israel

Received 5 March 2009; received in revised form 12 April 2009; accepted 14 April 2009.

Abstract 

The purpose of this study is to define the relationship between SPECT and CTA measured parameters of left ventricular (LV) function and volumes obtained in a single session using SPECT/64-slice CT hybrid imaging device, and in addition, to assess the reproducibility of LV parameters measured using 64-slice CTA.

Materials and methods

Seventy-six patients with suspected or known coronary artery disease underwent cardiac CTA and GSPECT in one session using a hybrid SPECT/CT device.

LV end-diastolic volume (EDV), end-systolic volume (ESV) and ejection fraction (EF) were measured on each component of the hybrid device. For the CTA component, these parameters were re-measured by the same investigator and by a second investigator with an interval of 3–54 weeks. Corresponding GSPECT and CTA measured parameters were compared. For CTA, intra-observer and inter-observer variability of LV function and volume measurements were calculated.

Results

A very good correlation was found between the GSPECT and CTA measured LVEF (r=0.81), ESV (r=0.90) and EDV (r=0.82). There was a small positive difference by CTA measured LVEF (3.9±14.2%), and more prominent positive differences by CTA measured ESV and EDV (9.8±14.8 and 44.9±23.1cm3, respectively). There was excellent reproducibility in the measurements of all parameters with very low intra- and inter-observer variability (r=0.93 for EF and 0.98 for EDV and ESV).

Conclusions

Although a good correlation was found between the EF measurements obtained from CTA and SPECT, interchangeable use of EF measurements between the two modalities should be done cautiously and interchangeable use of LV EDV and ESV should be avoided.

Keywords: GSPECT, Cardiac CT, Left ventricular function

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0720-048X(09)00230-7

doi:10.1016/j.ejrad.2009.04.039

European Journal of Radiology
Volume 75, Issue 2 , Pages 154-158, August 2010